All vår början bliver svår… Extern closed loop Ragnar Hanås, Barn- och Ungdomskliniken, Uddevalla och NÄL MAT ➠ Den första insulinpumpen från Indiana, USA 1963. ➠ Pumpen kunde även ge glukagon! Håller den blodsockerbalansen automatiskt? Low Glucose Suspend i praktiken Pump, sensor + LGS (Low Glucose Suspend 21 pat. 1-18 år, diabetes duration ≥ 12 mån., CSII 3 ≥ mån. 2 veckor med Sensor-Augmented Pump (SAP) jämfördes med 6 veckor med SAP + Low Glucose suspend (LGS) Hypoglykemi-larm vid 4,2 mmol/l, LGS vid 3,9 mmol/l Hannes, 8 år, 29 kg Diabetes sedan1½ år Pump, 0.7 E/kg/dag HbA1c 41 mmol/l (4,9%) Frukost-problem: Hyper-, sedan hypoglykemi Pumpen stänger av med LGS (Low Glucose Suspend) Resultat ➠ Hypo < 4.2 Hypo < 2.2 SAP SAP+LGS 1.27/dag 0.28/dag 0.95/dag 0.13/dag Tid i hypo 101 min/dag 58 min/dag < 4.2mmol/l Medel-BG 8.1 mmol/l 8.2 mmol/l Danne, T. Diab Tech Ther 2011;13, 1129-34. Danne, T. Diab Tech Ther 2011;13, 1129-34. Pump, sensor + LGS (Low Glucose Suspend) Closing the loop - InfuLink 50 pat. 17-58 years, diabetes duration > 1 year., CSII 3 ≥ months. Exercise sessions until BG < 4.7 mmol/l LGS set at 3.9 mmol/l, stopped basal for 120 min. 134 sessions, 36 not included as BG > 16.7 or < 2.8 mmol/l 5 mmol/l Exercise until 4.7 mmol/l 5 mmol/l 3.9 mmol/l Produkten finns inte! Children with Diabetes website Garg S. DiabTechTher 2012;14:205-09 1 Closed loop algoritmer S.c. sensor och s.c. pump hos barn och ungdomar 17 tonåringar med typ 1 diabetes, ålder 13-20 år MiniMed ePID pump och CGMS sänder radiosignaler till en laptop ➠ PID – proportional integral derivative - används mycket i kontrollsystem inom industrin - P: justerar i förhållande till aktuella PG-nivåer - I: arean under kurvan mellan uppmätt och target PG - D: hastigheten i PG-förändringen över tid 8 patienter fick allt insulin automatiskt enligt algorithm (FCL). 9 fick en liten manuell bolus före måltider (HCL). mmol/l 16,7 ➠ MPC – model predictive control - bättre på fördröjningar associerade med insulin-absorption - inkluderar måltider och måltidsdoser ➠ Fuzzy logic - baserad på läkarens sätt att ta beslut (“fuzzy logic“) - bygger på tidigare individuella patientdata - använder inte direkta fysiologiska modeller eller linjära 11,0 5,6 Medelvärde för BG på dagen var 8,3 ± 2,6 i HCL gruppen vs. 8,8 ± 3,3 i FCL gruppen (P = 0.03) matematiska beräkningar Hypoglykemi (<3,3 mmol/l) inträffade x1 i FCL & x2 HCL gruppen Kumareswaran K. Discov Med. 2012 Feb;13(69):159-70. Weinzimer SA. Diabetes Care. 2008; 31:934-9. Overnight closed loop in young children Overnight closed loop in adults 8 children, 6-13 years 77g carb. at 18:00h and 21g carb. at 21:00 (manual bolus) Navigator sensor + Deltec pump MPC algorithm, insulin dose adjusted automatically Results 12 adults consumed a medium sized meal (60g carb.) 12 adults consumed a larger meal (100g carb.) + wine (5 dl) Navigator sensor + Deltec pump MPC algorithm, insulin dose adjusted manually by nurse. Results ➠ Closed loop time in target (3.9-8.0 mmol/l) was 52% if started at 18:00 and 70% if stared at 21:00 (p=n.s.) ➠ Closed loop increased time in target (3.9-8.0 mmol/l) by 15% in medium meal and 28% in large meal ➠ Time < 3.9 mmol/l was 0% vs. 8% (p=n.s.) ➠ Closed loop decreased time in hypoglycemia (≤ 3.9 mmol/l) and eliminated time < 3.0 ➠ No rescue carbs for mmol/l after midnight hypoglycemia were needed Elleri D. DiabTechTher 2011;13:419-24. Hovorka R. BMJ 2011;342:d1855. Meta-analysis of Cambridge closed loop studies 4 randomized crossover studies in 17 children and adolescents and 24 adults (37.5 +/- 9.1 years) on 45 closed-loop and 45 CSII visits PG 3.9-8 mmol/l ➠ Results Time in target Time < 3.9 SD of PG CL CSII 60% 2.1% 1.5 40% 0.002 4.1% 0.01 2.1 0.07 100g carb. meal + 5 dl wine 60g carb. meal p Closed loop based on in silico design Algorithms based on in silico experiments (computer simulation), n=300 Clinical experiments with 20 adults from France, USA and Italy 22h of closed loop compared with 22h of open loop (dosing by patient under a physician’s supervision) Results ➠ Less nocturnal hypoglycemia PG PG>8.0 < 3.9 mmol/l mmol/l ➠ More time in target (3.9-7.8 mmol/l) ➠ No change during meal control Kumareswaran K. J Diabetes Sci Technol 2011;5:1352-62. Kovatchev B. J Diabetes Sci Technol 2010;4:1374-81. 2 Overnight closed loop in adolescents Att minimera riskerna Pilot study with 2 adolescents 2 Enlite sensors, Medtronic Veo pump Blackberry with Proportional-Integrative-Derivative (PID) ISM-to-Bluetooth radiofrequency (RF) translator SG #1 Results ➠ No investigator intervention Glucose (mg/dL) 280 ➠ Sensor glucose 3.4-8.4 mmol/l mmol/l SG (Control) Day (OL) MBG Night #2 (CL) 240 10 mmol/l 200 160 120 80 40 0 ePID Infusion Rate (U/h) ➠ YSI glucose 3.4-7.8 SG #2 Night #1 (CL) 5 4 3 2 1 0 20 22 24 26 28 30 32 34 36 38 40 42 44 46 48 50 52 54 20 22 24 26 28 30 32 34 36 38 40 42 44 46 48 50 52 54 Time (h) O’Grady MJ. ATTD 2012 Abstract. Closed loop in adults with fuzzy logic ➠ Returned to < 10 mmol/l (180 mg/dl) within 2.6 hours % 3.9-10.0 PID 73 MPC 79 EMPC 84 mmol/l mean BG 8.4 8.7 8.2 BB 7.8 92 Too high Extra insulin given Insulin dose given by algorithm (mean at minimal risk) Don´t give too much insulin quickly! Span of ± 75 mg/dl (4.2 mmol/l) Too low Wait for endogenous glucose production (takes much longer time to correct) Cameron F. J Diabetes Sci Technol 2011;5(2):368-379. Bi-hormonal closed loop USA Fuzzy logic theory to imitate the line of reasoning of diabetes caregivers 7 adults in Israel, 14 closed loop sessions of 8 and 24 hours Dexcom or Navigator sensors, Omnipod pumps Results ➠ Postprandial peak 12.5 mmol/l (224 mg/dl) Asymmetrisk risk för kliniska komplikationer med höga och låga BG Irreversibel verkan av insulin när bara insulin används Modellen testad i en silico simulator PID = proportional-integral-derivative MPC = model predictive control EMPC = enhanced MPC BB = basal/bolus controller (retrospective) Results 1 case report 2 Medtronic sensors, 2 Omnipod pumps and a laptop At calibration, the sensor that is most correct will be used until next calibration. Example patient 10 mmol/l Bolus dose ➠ 73% of sensor values 3.9-10 mmol/l (70-180 Basal insulin mg/dl), 0% < 3.9 Atlas E. Diabetes Care 2010;33:1072-6. Bi-hormonal closed loop USA 11 patients 2 pumps and a laptop, IV BG every 5 min. 27 hour experiments with 3 meals. No premeal announcements. MPC algorithm with PK model of lispro, PK model for glucagon Jacobs_Conf Proc IEEE Eng Med Biol Soc. 2011;2011.397-400. Bi-hormonal closed loop USA 10 mmol/l 10 mmol/l 3.9 mmol/l 2.82.8 mmol/l mmol/l 10 10mmol/l mmol/l 3.9 mmol/l 3.9mmol/l mmol/l 3.9 2.8 mmol/l 2.8 mmol/l Results ➠ Time to peak 117 min. in group with hypoglycemia vs. 64 min with no hypo ➠ Time in algorithm was set to 31 min! No hypoglycemia treatment needed adjusted settings lispro in same patients! Hypoglycemia treatment needed withwith standard PKPK settings for for lispro! ➠ Better results with no hypo in either group after adjustment of lispro PK to give peak after 65 min. El-Khatib FH. Sci Transl Med 2, 27ra27 (2010) 10 mmol/l El-Khatib FH. Sci Transl Med 2, 27ra27 (2010) 3 Uppkopplad i bilen…! Future product concept not currently available for investigational use. Has not been reviewed by the U.S. FDA. 4